Testomat 2000® PO4
Product number: 100560
8.047,80 €*
* Prices excl. VAT plus shipping costs.
Product information
Dimensions: 380 x 280 x 480 mm
Weight: 10,600 kg
Measurement parameters: Phosphate PO4
Measuring range: 0.0 – 10.0 mg/l Resolution: 0.1 mg/l in the range of 0.0 – 7 mg/l 0.25 mg/l in the range of 7 – 10 mg/l Limit of detection: 0,1 or 0,25 mg/l depending on range Accuracy: < 10% at end of measuring range
Range of application: treated wastewater (sewage treatment plants) | online – environmental analysis | monitoring of conditioning agents in cooling and heating circuits
Menu navigation: German, English, French, Dutch, Spanish
Description
Description "Testomat 2000® PO4"
The Testomat 2000® PO4 process photometer is an online analyzer for monitoring the content of ortho – phosphate in the range 0 to 10.0 mg/l (ppm) by using the photometric analysis principle.Performance profile:
- Analysis via automatic dosing of two reagents
- The analysis result is displayed after a reaction time of approx. 10 min. (Measuring time without rinsing time)
- Menu-driven operation and programming via plain text display
- Selection of unit in ppm or mg/l
- High-precision titration using a piston dosing pump
- Analysis initiation:
- Automatic interval operation (Interval pause adjustable from 0–255 min)
- External control
- Dynamic (fatigue-dependent interval operation)
- Quantity dependent via contact water meter
- Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
- Monitoring of two measuring points (Switching by external solenoid valves)
- Programmable service address
- Internal error documentation
- Programmable maintenance interval for maintenance request