Product information

Dimensions: 380 x 280 x 480 mm
Weight: 10,600 kg
Measurement parameters: Free chlorine
Measuring range: 0.00 – 0.99 mg/l (Resolution: 0.01 mg/l) 1 – 2.5 mg/l (Resolution: 0.1 mg/l)
Range of application: Chlorination monitoring for drinking-water or swimming pools. | Provides protection for reverse osmosis membranes. | Control of chlorine biocides and conditioning agents.
Menu navigation: German, English, French

Description

Description "Testomat 2000® CLF"

The Testomat 2000® CLF process photometer is an online analyzer for monitoring the
 ontent of free chlorine (CLF) in the range of 0-2.5 mg / L (ppm).

Performance profile:
  • Analysis via automatic dosing of two reagents
  • The analysis result is displayed after a reaction time of approx. 1 min. (Measuring time without rinsing time)
  • 2 independently programmable limit contacts for monitoring and control tasks
  • Menu-driven operation and programming via plain text display
  • High-precision titration using a piston dosing pump
  • Analysis initiation:
    • Automatic interval operation
    • (Interval pause adjustable from 0–99 min)
    • External control
    • Dynamic (fatigue-dependent interval operation)
    • Quantity dependent via contact water meter
  • Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
  • Monitoring of two measuring points (Switching by external solenoid valves)
  • programmable service address
  • Internal error documentation
  • Programmable maintenance interval for maintenance request