Testomat 2000®

Product number: 100011
4.662,70 €*
* Prices excl. VAT plus shipping costs.
Heyl

Available variants

Language in the display: Dutch, English, French, German, Italian, Polish, Spanish
Performance/execution: 115 V/50-60 HZ, 230 V/50-60 HZ, 24 V/50-60 HZ

Product information

Dimensions: 380 x 280 x 480 mm
Weight: 10,600 kg
Measurement parameters: Water hardness | Carbonate hardness | p-value | minus m-value
Measuring range: 0.89 - 448 ppm | 8.9 - 358 ppm | 0.1 - 15 mmol/l | 0.05 - 0.5 mmol/l
Range of application: Universal for water treatment plants
Menu navigation: German, English, French, Polish, Dutch, Italian, Spanish

Description

Description "Testomat 2000®"

The process photometer Testomat 2000® automatically determines water hardness, carbonate hardness, the p-value or the minus m-value via titration. It is ideal for monitoring the water quality of water treatment, water blending and drinking-water plants. Monitoring and quality control of softening plants with and without blending (Testomat 2000®  reagent TH 2005–2250)
Monitoring and quality control of decarbonisation plants (Testomat 2000® reagent TC 2050–2100) Measuring acidic values in decarbonisation and desalination plants (Testomat 2000® reagent TM 2005) Monitoring of steam boilers (Testomat 2000® reagent TP 2100).

Performance profile:
  • Menu-driven operation and programming via plain text display
  • Determinal measuring ranges of residual hardness, total hardness, carbonate hardness, minus m value, p value by indicator selection
  • Selection of hardness units in ° dH, ° f, ppm CaCO3 or mmol / l
  • High-precision titration using a piston dosing pump
  • Analysis initiation:
    • - Automatic interval operation (Interval pause adjustable from 0–99 min)
    • - External control
    • - Dynamic (fatigue-dependent interval operation)
    • - Quantity dependent via contact water meter
  • Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
  • Monitoring of two measuring points (Switching by external solenoid valves)
  • Internal error documentation
  • Programmable service address
  • Indicator quantities: 100 and 500 ml bottles
  • Programmable maintenance interval for maintenance request

Shop